Radiation tolerance of epitaxial silicon carbide detectors for electrons, protons and gamma-rays
Articolo
Data di Pubblicazione:
2003
Abstract:
Particle detectors were made using semiconductor epitaxial 4H–SiC as the detection medium. The investigated
detectors are formed by Schottky contact (Au) on the epitaxial layer and an ohmic contact on the back side of 4H–SiC
substrates with different micropipe densities from CREE. For radiation hardness studies, the detectors have been
irradiated with protons (24 GeV=c) at a fluence of about 1014 cm2 and with electrons ð8:2 MeVÞ and gamma-rays
(60Co source) at doses ranging from 0 to 40 Mrad: We present experimental data on the charge collection properties by
using 5.48, 4.14 and 2:00 MeV a-particles impinging on the Schottky contact. Hundred percent charge collection
efficiency (CCE) is demonstrated for reverse voltages higher than the one needed to have a depletion region equal to the
a-particle projected range, even after the irradiation at the highest dose. By comparing measured CCE values with the
outcomes of drift–diffusion simulations, values are inferred for the hole lifetime, tp; within the neutral region of the
charge carrier generation layer. tp was found to decrease with increasing radiation levels, ranging from 300 ns in nonirradiated
detectors to 3 ns in the most irradiated ones. The diffusion contribution of the minority charge carriers to
CCE is pointed out.
Tipologia CRIS:
03A-Articolo su Rivista
Keywords:
Radiation hardness; Device simulation; Silicon carbide; Semiconductor detectors; Charge–particle spectroscopy
Elenco autori:
F. NAVA; E. VITTONE; P. VANNI; G. VERZELLESI; P.G. FUOCHI; C. LANZIERI; M. GLASER