Data di Pubblicazione:
2006
Abstract:
We present a quick and easy method for the calibration of a matrix of sensors. The algorithm is based on a three-step irradiation procedure which relies only on the constancy of the delivered fluence at each step. With this method the gain of each sensor is derived relative to a reference detector. The algorithm has been applied to a matrix of (32 x 32) ionization chambers. After the calibration coefficients have been applied, by comparing the response of the matrix of chambers to a reference detector over a large field, we determined that the fluence measurement of individual chambers is better than 0.7%. The algorithm solves the cumbersome problem of the relative gain calibration of a matrix of a large number of sensors.
Tipologia CRIS:
03A-Articolo su Rivista
Elenco autori:
M. DONETTI; E. GARELLI; F. MARCHETTO; A. BORIANO; F. BOURHALEB; R. CIRIO; I. CORNELIUS; S. GIORDANENGO; A. LA ROSA; U. NASTASI; C. PERONI